The illuminance responsivity determines how accurately the meter responds to the light level. The spectral responsivity determines how accurately the colour response of the meter matches that of the eye. Both are calibrated against our reference detectors.
The meter is calibrated against a reference illuminance detector using a light source, typically incandescent Illuminant A or an LED source, to determine its performance.
To determine the spectral responsivity, light from an Illuminant A source is passed through four filters in turn: red, green, blue and infrared. The meter response is referenced to the response of an ideal detector. The results are recorded in a NATA-endorsed report.
As a general rule meters should be recalibrated every year. If a meter is to be used in a legal case it should have been recalibrated within the preceding 12 months.
The results of the calibration, including the meter response to the filter set, the IR response and the spectral response.
The laboratory maintains reference detectors that are traceable to the SI system of units, so your readings can be defended.
Compared with the previous report, each calibration shows how stable the meter is over time and whether it has drifted or been damaged.