HomeCalibrationLux & Luminance Meter Calibration
Calibration · Lux and luminance meters

Lux and luminance meter calibration.

As for any electronic measuring device, a lux meter needs to be checked against a known reference to make sure it is reading accurately. A calibrated meter yields accurate and consistent measurements over many years.

Illuminance
0.1–10,000 lux
Luminance
0.1–30,000 cd/m²
Source
Illuminant A · LED
Accreditation
NATA 2258
What we calibrate

The two things that decide whether a meter reads true.

The illuminance responsivity determines how accurately the meter responds to the light level. The spectral responsivity determines how accurately the colour response of the meter matches that of the eye. Both are calibrated against our reference detectors.

Illuminance responsivity
Lux meters
Luminance responsivity
Luminance meters
Spectral responsivity
Red, green, blue, IR filters
Linearity
Across the range
Cosine response
Optional
Colour meters
On request
In detail

How a lux meter is calibrated.

The meter is calibrated against a reference illuminance detector using a light source, typically incandescent Illuminant A or an LED source, to determine its performance.

To determine the spectral responsivity, light from an Illuminant A source is passed through four filters in turn: red, green, blue and infrared. The meter response is referenced to the response of an ideal detector. The results are recorded in a NATA-endorsed report.

As a general rule meters should be recalibrated every year. If a meter is to be used in a legal case it should have been recalibrated within the preceding 12 months.

What you receive

A report you can file and rely on.

NATA-endorsed report

The results of the calibration, including the meter response to the filter set, the IR response and the spectral response.

Traceable results

The laboratory maintains reference detectors that are traceable to the SI system of units, so your readings can be defended.

A calibration history

Compared with the previous report, each calibration shows how stable the meter is over time and whether it has drifted or been damaged.

Frequently asked

Common questions, answered.

As a general rule, every year. If readings may be used in a legal matter, the meter should have been calibrated within the preceding 12 months, and it is good practice to check it just before or just after the survey.
Ready to begin

Send us your sample,
we’ll send back certainty.

Request a quote for testing or calibration and hear back from a technician, not a sales representative, within one business day.